Key Specifications
| Description |
Liquid-to-Liquid Cooler |
|---|---|
| Connectivity |
Optional: RS232, RS485, Remote On/Off
|
| Working Temperature Range °C |
Facility water +10° to 60°
|
| Temperature Stability °C |
±0.4°
|
| Temperature Calibration Capability |
1-point
|
| Pump Type |
Turbine
|
| External Temperature Control Capability |
Optional
|
Key Features
- Protects precision equipment from facility water contaminants
- Quiet, energy-efficient liquid-to-liquid heat removal
- Displays temperature and pressure or flow rate information
- Built-in temperature and low flow alarms
Related Fluids & Accessories
Common Specifications
| Working Temperature Range °F |
Facility water +20° to 140°
|
|---|---|
| Working Temperature Range °C |
Facility water +10° to 60°
|
| Reservoir Capacity (gallons) |
1.11
|
| Reservoir Capacity (liters) |
4.2
|
| Reservoir Cover |
Screw-on Cap
|
| Cleanable Fluid Filter |
Yes
|
| Temperature Stability °F |
±0.8°
|
| Temperature Stability °C |
±0.4°
|
| Pump Type |
Turbine
|
| Pump Speed |
Constant, Single Speed
|
| Process Connections |
1/2" (F) NPT
|
| External Temperature Control Capability |
Optional
|
| Over-Temperature Protection / Failsafe Heater Control |
Yes
|
| Reservoir Drain |
Yes
|
| Maximum Ambient Temperature °F |
95°
|
| Maximum Ambient Temperature °C |
35°
|
| Overall Dimensions (L x W x H) (inches) |
27.6 x 14.5 x 22.6 in
|
| Overall Dimensions (L x W x H) (cm) |
70.2 x 36.8 x 57.5 cm
|
| Shipping Weight (pounds) |
175.0
|
| Shipping Weight (kilograms) |
79.5
|
| Catalog Page Number |
108
|
60 Hz Only
| Part Number |
4150T21A130D
|
|---|---|
| Maximum Pressure (psi) |
100.0
|
| Maximum Pressure (bar) | |
| Maximum Pressure Flow Rate (gpm) |
3.50
|
| Maximum Pressure Flow Rate (l/min) |
13.2
|
| Cooling Capacity (W/BTU) |
15000 / 52855 based on 15°C ΔT
10000 / 34100 based on 10°C ΔT
4500 / 5345 based on 5°C ΔT
ΔT = Process water temperature - facility water temperature
|
| Included Hardware |
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
|
| Electrical Requirements (VAC/Hz/Ph/A) |
208-230/60/1/3
|
| Regulatory Approvals |
CSA
|
50 Hz Only
| Part Number |
4150T21A130E
|
|---|---|
| Maximum Pressure (psi) |
83.0
|
| Maximum Pressure (bar) | |
| Maximum Pressure Flow Rate (gpm) |
2.90
|
| Maximum Pressure Flow Rate (l/min) |
11.0
|
| Cooling Capacity (W/BTU) |
15000 / 52855 based on 15°C ΔT
10000 / 34100 based on 10°C ΔT
4500 / 5345 based on 5°C ΔT
ΔT = Process water temperature - facility water temperature
|
| Included Hardware |
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
|
| Electrical Requirements (VAC/Hz/Ph/A) |
240/50/1/3
|
| Regulatory Approvals |
CE
|