Key Specifications
Description |
Immerson Probe-60, 1.5" Rigid Coil |
---|---|
Working Temperature Range °C |
-60° to -20°
|
Temperature Control (°C) |
Fixed at -60°
|
Key Features
- Continuous cooling to temperatures as low as -100°C
- Designed to run at maximum cooling
- An economical alternative to dry ice or liquid nitrogen
- Excellent for trapping applications, freeze drying, and rapidly cooling small volumes of liquids
Common Specifications
Working Temperature Range °F |
-76° to -4°
|
---|---|
Working Temperature Range °C |
-60° to -20°
|
Temperature Control (°F) |
Fixed at -76°
|
Temperature Control (°C) |
Fixed at -60°
|
Maximum Ambient Temperature °F |
86°
|
Maximum Ambient Temperature °C |
30°
|
Overall Dimensions (L x W x H) (inches) |
11 x 10 x 9 in
|
Overall Dimensions (L x W x H) (cm) |
27.9 x 25.4 x 22.9 cm
|
Refrigerant |
R404A & R508B
|
Immersion Probe Type |
Rigid Coil
|
Probe Diameter (inches) |
1.5
|
Probe Diameter (cm) |
3.81
|
Probe Length (inches) |
Coil: 4 / Exposed: 15
|
Probe Length (cm) |
Coil: 10.2 / Exposed: 38.1
|
Flexible Hose Diameter (inches) |
1.5
|
Flexible Hose Diameter (cm) |
3.81
|
Flexible Hose Length (feet) |
4
|
Flexible Hose Length (m) |
1.22
|
Shipping Weight (pounds) |
49.0
|
Shipping Weight (kilograms) |
22.0
|
Catalog Page Number |
114-115
|
60 Hz Only
Part Number |
P60N2A101B
|
---|---|
Cooling Capacity @ -60°C (W) |
0
|
Cooling Capacity @ -40°C (W) |
100
|
Cooling Capacity @ -30°C (W) | 125 |
Cooling Capacity @ -20°C (W) |
150
|
Electrical Requirements (VAC/Hz/Ph/A) |
120/60/1/5
|
50 Hz Only
Part Number |
P60N2A102E
|
---|---|
Cooling Capacity @ -60°C (W) |
0
|
Cooling Capacity @ -30°C (W) | 104 |
Cooling Capacity @ -20°C (W) |
125
|
Electrical Requirements (VAC/Hz/Ph/A) |
240/50/1/2.5
|
Regulatory Approvals |
CE
|