Key Specifications
Description |
6200 Series 1/4 HP Chiller, Centrifugal Pump; 10 psi; 15.5 l/min |
---|---|
Display |
LED
|
Connectivity |
Optional: RS232, RS485, Remote On/Off
|
Working Temperature Range °C |
-10° to 40°
|
Temperature Stability °C |
±0.1°
|
Temperature Calibration Capability |
1-point
|
Pump Type |
Centrifugal
|
External Temperature Control Capability |
Optional
|
Key Features
- Large, dual displays present temperature and pressure or flow rate simultaneously
- Compact, portable design takes up less floor space
- Cooling at ambient temperatures as high as 35°C
- Magnetic drive centrifugal pump
- Choice of compressor sizes
- User-adjustable temperature, pressure, and flow rate alarms
- External temperature tracking and communications capability (optional)
- Cool Command™, WhisperCool® (3/4 and 1 HP models)
Related Fluids & Accessories
Common Specifications
Working Temperature Range °F |
14° to 104°
|
---|---|
Working Temperature Range °C |
-10° to 40°
|
Reservoir Capacity (gallons) |
1.11
|
Reservoir Capacity (liters) |
4.2
|
Reservoir Cover |
Screw-on Cap
|
Cleanable Air Filter |
Yes
|
Cleanable Fluid Filter |
Yes
|
Temperature Stability °F |
±0.2°
|
Temperature Stability °C |
±0.1°
|
Display |
LED
|
Pump Type |
Centrifugal
|
Pump Speed |
Constant, Single Speed
|
Process Connections |
1/2" (F) NPT
|
External Temperature Control Capability |
Optional
|
Over-Temperature Protection / Failsafe Heater Control |
Yes
|
Reservoir Liquid Level Indicator |
Yes
|
Cool Command Refrigeration Technology |
Yes
|
Reservoir Drain |
Yes
|
Heater Option |
Yes
|
Maximum Ambient Temperature °F |
95°
|
Maximum Ambient Temperature °C |
35°
|
Overall Dimensions (L x W x H) (inches) |
27.6 x 14.5 x 22.6 in
|
Overall Dimensions (L x W x H) (cm) |
70.2 x 36.8 x 57.5 cm
|
Compressor Size (HP) |
1/4
|
Refrigerant |
R134a
|
Shipping Weight (pounds) |
131.0
|
Shipping Weight (kilograms) |
59.4
|
Catalog Page Number |
102-103
|
60 Hz Only
Part Number |
6260M11A110B
|
---|---|
Maximum Pressure (psi) |
10.0
|
Maximum Pressure (bar) | |
Maximum Pressure Flow Rate (gpm) |
4.10
|
Maximum Pressure Flow Rate (l/min) |
15.5
|
Cooling Capacity @ -10°C (W) | 150 |
Cooling Capacity @ -5°C (W) |
225
|
Cooling Capacity @ 0°C (W) | 300 |
Cooling Capacity @ +10°C (W) | 700 |
Cooling Capacity @ +20°C (W) |
950
|
Cooling Capacity @ +30°C (W) | 1125 |
Cooling Capacity @ +40°C (W) | 1275 |
Included Hardware |
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
|
Electrical Requirements (VAC/Hz/Ph/A) |
120/60/1/9.5
|
Regulatory Approvals |
TUV
|
50 Hz Only
Part Number |
6250M21A130E
|
---|---|
Maximum Pressure (psi) |
9.5
|
Maximum Pressure (bar) | |
Maximum Pressure Flow Rate (gpm) |
3.90
|
Maximum Pressure Flow Rate (l/min) |
14.7
|
Cooling Capacity @ -10°C (W) | 125 |
Cooling Capacity @ -5°C (W) | 260 |
Cooling Capacity @ 0°C (W) |
400
|
Cooling Capacity @ +10°C (W) | 600 |
Cooling Capacity @ +20°C (W) |
800
|
Cooling Capacity @ +30°C (W) | 1050 |
Cooling Capacity @ +40°C (W) | 1125 |
Included Hardware |
Two sets of Inlet/Outlet Adapters: ½ inch male NPT, 5/8 inch male NPT
|
Electrical Requirements (VAC/Hz/Ph/A) |
240/50/1/5.6
|
Regulatory Approvals |
CE
|