Product Specification |
Description | |
pH | ||
Range | -2.00 to 16.00 pH | |
Resolution | 0.01 pH | |
Accuracy | ±0.01 pH | |
ORP | ||
Range | -1000 to 1000 mV / 0 to 100.0 % | |
Resolution | 1 mV / 0.1 % | |
Accuracy | ±1 mV / ±0.2 % | |
Temperature | ||
Range | -9.9 to 125.0 ºC | |
Resolution | 0.1 ºC | |
Accuracy | ±0.5 ºC | |
Sensor | Pt100 / Pt1000 (jumper selectable); 2 or 3 wire | |
Compensation | Auto/manual | |
Set Point And Controller Functions | ||
Set point 1 (SP1) / set point 2 (SP2) |
0.00 to 14.00 pH or -1000 to 1000 mV or 0 to 100 % |
|
Switching pH Hysteresis | 0.1 to 1 pH | |
Switching ORP Hysteresis | 10 to 100 mV / 1 to 10.0 % | |
Function (Switchable) | P control (pulse length/pulse frequency); limit control |
|
Adjustable Period With Pulse Length Controller | 0.5 to 20 sec | |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min | |
Pickup/Dropout Delay | 0 to 2000 sec | |
Contact Outputs | 3 SPDT relays | |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA | |
Alarm Functions | ||
Function (Switchable) | Steady or fleet (pulse) | |
Wash Cycle | 0.1 to 199.9 hr | |
Wash Direction | 1 to 1999 sec | |
Pickup Delay | 0 to 2000 sec | |
Switching Voltage/ |
Max. 250 VAC / max. 3 A / max. 600 VA | |
Electrical Data And Connections | ||
Transmitter Function | 0/4 to 20 mA scalable outputs for pH/ORP, galvanically isolated | |
Hold Function Switch | To freeze output current and deactivate control relays | |
Load | Max. 600 Ohm | |
pH/ORP Input | BNC (1012 impedance); asymmetrical/symmetrical | |
Connection Terminal | 5-pole, 17-pole terminal, detachable blocks | |
Display | ||
LCD | 7 segments display with symbols for status information | |
Power Supply | ||
Input | 110 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSCTP1001) |
220 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSCTP1002) |
Main Fuse | Slow-blow 250 V / 100 mA | |
Pollution Degree | 2 | |
Transient Overvoltage Category | II | |
EMC Specifications | ||
Emitted Interference | According to EN 50081-1 | |
Immunity To Interference | According to EN 50082-1 | |
Environmental Conditions | ||
Operating Temperature Range | -10 to 50 ºC | |
Max. Relative Humidity | 80 % up to 31 ºC decreasing linearly to 50 % at 40 ºC |
|
Mechanical Specifications | ||
Dimensions (WxHxD) | 96 x 96 x 175 mm | |
Weight | 700 g (unit) / 800 g (packed) | |
Ingress Protection | IP54 (front panel) |