Your Water Testing Needs, All in One Spectrophotometer

The DR 6000 has the most pre-programmed testing methods, including high-speed wavelength scanning across the UV and Visible Spectrum.

Accessories for High Volume and High Accuracy Testing Needs

A carousel sample changer allows up to seven sequential measurements. The Sipper Module, an instrument-controlled sample delivery system, increases precision by constant optical characteristics.

Advanced Quality Assurance at Your Fingertips

The DR 6000 comes with integrated QA software for scheduling, documenting and interpreting all of your needed quality measurements.

Guided Procedures and Elimination of False Readings

The DR 6000, when used with TNTplus® reagent vials, provides the accurate results you need by guiding you step-by-step through your testing procedures. With TNTplus, the instrument averages 10 readings and eliminates outliers, making scratched, flawed or dirty glassware a non-issue.

Automatically Avoids Errors

RFID* technology automatically updates the program calibration factors when you place a TNTplus reagent box near the DR 6000. The instrument identifies chemistry expiration dates via a barcode on the vials, and detects chemistry coefficient factors to avoid errors that can occur in lot-to-lot variations in the chemistry. 
*RFID technology currently available only in US, Anguilla, American Samoa, Australia, Bolivia, Canada, Cayman Islands, Columbia, Dominican Republic, El Salvador, Federated States of Micronesia, Guam, Guatemala, Marshall Islands, New Zealand, Northern Mariana Islands, Palau, Panama, Puerto Rico, and US Virgin Islands.

  What's in the box?

Includes: printed instrument manual, power cords (115V and 230V), universal holder, dust cover, 1-in. square matched glass sample cells, and a universal-adapter

  Specifications

Data Logger:5000 data points (result, date, time, sample-ID, user-ID)
Dimensions (H x W x D):215 mm x 500 mm x 460 mm
Display:TFT 7 inch WVGA color touch
Enclosure Rating:IP20 with closed lid
Includes:1 x Power Cord (US, EU)
1x Universal-Adapter
1x Dust Cover
Matched pair of 1 inch glass sample cells
Printed multilingual basic user manaul (en, es, fr, ja, ko, pt, zh)
Length:460 mm
Manual Languages:en, es, fr, ja, ko, pt, zh
Operating Conditions:10 to 40 °C, max. 80% relative humidity (non-condensing)
Operating Mode:Transmittance (%), absorbance and concentration (wavelength, time)
Optical System:Reference beam, spectral
Reference beam, spectral
Photometric Accuracy:5 mAbs at 0.0 to 0.5 Abs
<1% at 0.5 to 2.0 Abs at 546 nm
Photometric Linearity:0.005 to 2 Abs
≤ 0.01 at > 2 Abs with neutral glass at 546 nm
Photometric Measuring Range:± 3 Abs
Power Requirements:100 - 240 V; 50/60 Hz
Preprogrammed Methods:> 240
Sample Cell Compatibility:Rectangular: 10, 20, 30, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch

Optional 100 mm rectangular cell with additional adapter
Scanning Speed:900 nm/min (in 1 nm steps)
Source Lamp:Tungsten (visible range), deuterium (UV range)
Specific Technology:RFID for easy method update, sample ID and Certificate of Analysis
Spectral Bandwidth:2 nm
Storage Conditions:-25 to 60 °C / max. 80% relative humidity (non-condensing)
Stray Light:KI-solution at 220 nm < 3.3 Abs/ < 0.05%
User Interface Languages:bg, cn, cz, da, en, es, fr, gr, hr, hu, it, jp, kr, nl, pl, pt, ro, ru, sl, sl, sv, tr
User Programs:200
Warranty:1 year
Wavelength Accuracy:± 1 nm
Wavelength Range:190 to 1100 nm
Wavelength Reproducibility:< 0.1 nm
Wavelength Resolution:0.1 nm
Wavelength Selection:Automatic, based on method selection
Weight:24.25 lbs. (11 kg)