Samples are applied to a scratch-resistant zinc selenide attenuated total reflectance (ATR) sample plate and easily cleaned with a cloth or paper towel. No solvents requiring special ventilation, expensive waste disposal or difficult to clean cells are needed. The analysis can be done by non-technical personnel with no formal training and is ideal for lab or field use. The InfraCal2 has added features such as touch screen display, data storage and transfer, a multi-language interface and password protected settings. The internal battery pack allows for portability and up to 12 hours continuous use, providing results in under a minute at a fuel loading rack, in a laboratory or at a pump in a fueling station.
- Measurement range: 0 - 100% ethanol
- Instrument repeatability: +/- 0.2
- Compares to ASTM D4815
Features & Benefits
Benefits |
Features |
Internal Data Storage |
Data storage including each measurement tagged with analyst, location and date/time information, and password protected instrument settings |
Export |
Directly to your PC via USB or a Flash Drive |
Touch Screen Display |
The touch screen display allows a simple set up for quick operation along with an easy calibration wizard. |
Internal Battery |
The internal battery pack allows for portability and up to 12 hours continuous use. |
Multi-Language Interface |
English, Spanish, French, Arabic |
Password Protected Settings |
Calibration and data are password protected to avoid accidental deletion |
Multiple Calibrations |
A Multitude of calibrations can be stored allowing the InfraCal to be moved to different locations without the need to recalibrate after initial calibration at a particular location |
No need for Installation/Commissioning |
The InfraCal 2 is a portable analyzer, weighing less than 5 lbs. with a 6 in. square foot print, requiring no installation or commissioning, little to no routine maintenance or recalibration. |
Minimal Consumables |
The analysis requires little to no consumables making the cost-per-test minimal. |
Highlights
The InfraCal 2 Model ATR-E can be ordered with a choice of one of both of the following factory calibrations. The two ranges of calibration available are 0-20% and 70-100%. A 0-20% 5 Point Ethanol Low Concentration Calibration includes 0, 5, 10, 15, and 20% ethanol in gasoline points. A 70-100% 5 Point Ethanol High Concentration Calibration includes 70, 80, 90, 95, and 100% ethanol in gasoline. The calibration is performed using standards prepared by a certified laboratory.
Specifications:
Instrument Type |
Fixed Filter Infrared Analyzer |
---|---|
Analytical Wavelength/Wave Number |
9.6 µm, 1045cm-1 |
Power Requirements |
18 volts DC, 3.3 amps, internal battery pack available |
Power Supply |
Universal AC/DC provided |
Weight |
5.8 lb. (2.6 kg) with battery -7.0 lbs (3.2 kg) |
Dimensions |
(17 cm) x 7.8" (19.8 cm) x 5.2" (13.2 cm) |
Suggested Operating Range |
5°C - 40°C |
Typical Useful Measurement Range |
0 - 98% Ethanol in Gasoline |
Analysis Time |
Less than 30 seconds |
Instrument Repeatability |
0 - 20%±0.2%, 20 - 98%±1% of ethanol in petrol concentration |
Communication Port |
USB, RS 232 |