Universal Oven UN160
Standard delivery Universal Oven
The universally applicable lab oven U is Memmert’s classic appliance for temperature control in science, research and material tests in industry. The technologically perfected masterpiece made of high-quality, hygienic, easy-to-clean stainless steel leaves nothing to be desired in terms of ventilation and control technology, overtemperature protection and precisely tuned heating technology.
On this page, you can find all the essential technical data on the universal Memmert lab oven. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.
Properties
Control of standard components
Temperature | 1 Pt100 sensor DIN class A in 4-wire-circuit |
---|---|
ControlCOCKPIT | adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Temperature
Set temperature range in °C | min. 5°C above ambient up to +300°C |
---|---|
resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED) |
Control technology
Function SetpointWAIT | the process time does not start until the set temperature is reached |
---|---|
Language setting | German/English/Spanish/French |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
Ventilation
natural convection | |
---|---|
Fresh air admixture | adjustment of pre-heated fresh air admixture by air flap control in 10 % steps |
Vent | vent connection with restrictor flap |
Communication
Documentation | programme stored in case of power failure |
---|---|
Programming | AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand). |
Safety
Temperature control | adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
---|---|
Autodiagnostic system | for fault analysis |
Standard equipment
Scope of delivery | incl. works calibration certificate for +160°C |
---|---|
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
Housing | rear zinc-plated steel |
Interior | easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Internals | 2 stainless steel grids |
Stainless steel interior
Max. loading of chamber: 210 kg | |
Dimensions W x H x D in mm | w(A) x h(B) x d(C): 560 x 720 x 400 mm |
---|---|
Volume 161 l |
Textured stainless steel casing
w(D) x h(E) x d(F): 745 x 1104 x 584 mm |
Electrical data
Voltage Electrical load | 230 V, 50/60 Hz approx. 3200 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 1800 W |
Packing/shipping data
the appliances must be transported upright | |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton B x H x T: 830 x 1300 x 800 mm | |
Net weight approx. 96 kg | |
Gross weight carton approx. 122 kg |